Release Level Date Purpose
R10V 09/08/2010 Update
R10U 06/21/2010 Update
R10T 06/04/2010 Update
R10R 03/10/2010 Update
R10Q    02/08/2010 Update
R10P 01/14/2010 Update
R10N 01/07/2010 Update
R10M 12/18/2009 Update
R10L 07/01/2009 Update
R10K 06/24/2009 Update
R10J 05/06/2009 Update
R10H 03/06/2009 Update
R10G 03/04/2009 Update
R10F 02/26/2009 Update
R10E 02/24/2009 Product Launch

R10V

EVM [All Models]

Updates

  • Added support for ozone (O3)
  • Added support for mechanical impactors.

Back to top of page

R10U

EVM [All Models]

Updates

  • PID (PPM sensor) displays hundredths from 0 to 500 mg/m^3 (was displaying tenths)

Back to top of page

R10T

EVM [All Models]

Updates

  • Measurements - PID (PPM sensor) measurement is displayed in hundredths from 0 to 20 PPM.
  • Calibration - Pump duty cycle is displayed while performing a user calibration. (Prior to this release, it could display dashes).

Back to top of page

R10R

EVM [All Models]

Updates

  • Added molecular weight to log files allowing QuestSuite to convert between PPB (PPM) and ug/m^3 (mg/m^3) for PID measurements (used for LEED initiative).

Back to top of page

R10Q

EVM [All Models]

Updates

  • Measurements and calculations- Added ug/m^3 as an alternate user unit for PID (PPB) sensor.
  • Measurements and calcuations- Added mg/m^3 as an alternate user unit for PID (PPM) sensor.

Back to top of page

R10P

EVM [All Models]

Updates

  • PID ppb (sensor) Internal Calibration Update: Changes the baseline (zero ppb) as requested by the manufacturer.

Back to top of page

R10N

EVM [All Models]

Updates

  • Language update: Revised Portuguese translations for PID, RH (Relative Humidity) and Dew Point.
  • Particulate Measurements: Corrected the occasional false positive "laser not working properly" error message on extended logs (4+ hours).
  • CO2 and PID measurements: Corrected the infrequent CO2 and PID measurement errors (CO2 measurement goes to zero, PID measurement goes to > 1 PPM).
  • Display: Various minor display issues resolved.

Back to top of page

R10M

EVM [All Models]

Product Enhancements

  • Auto-Run with Date, Date of Week, and Timed-Run modes have been programmed, tested and released.
  • File System Sessions menu: Added sorting to file system sessions. Sessions are now displayed in date descending order. (This is accessed from the following menus: From Start screen: File System/Session Directory. Press Enter to toggle through 3 different views.)

Updates

  • PID measurements: PID lamp is turned off when not needed to extend life.
  • PID PPB calibration: PID (PPB only) sensor only supports a span calibration (user zero calibrations are not needed).

Back to top of page

R10L

EMV [All Models]

Displayed System Errors

  • Eliminated multiple user notifications when the file system is full during a logging session.
  • Eliminated multiple user notifications when the system is in secure run mode and the impactor position is changed during logging.

Back to top of page

R10K

EVM [All Models]

Updates

Measurements and calculations

  • Allow users the option to save additional log interval values: Lmax, Lmin, Lavg and Lmax_stel.
  • Added dew point as a (calculated) parameter.
  • Added ug/m3 as an alternate user unit for particulate matter (PM).

PID Measurements

  • Added a PhotoIonization (PID) correction factor for the PID sensor when measuring a specific Volatile Organic Compound (VOC).
  • Added support for a Parts Per Billion (PPB) Photoioniztion (PID) sensor.

Security/Lock feature Added

  • Added secure run and setup modes.

Back to top of page  

R10J

EVM [All Models]

Updates

Particulate Matter Pump

  • Disable the Particulate Matter (PM) pump when it is unable to regulate at 1.67 L/min.

Calibration

  • The calibration screens now displays the details of factory calibrations.
  • The user settable span calibration value is set to the previous user span calibration value as a convenience.

Unit Info screen updated

  • The hardware version is now displayed in the UNIT INFO screen.

Back to top of page

R10H

EVM [All Models]

Updates

CO2 sensor

  • Corrected a timing issue that caused the CO2 sensor value to go under range periodically.
Back to top of page
 
R10G

EVM [All Models]

Updates

  • Corrected a hardware issue sometimes leading the unit to display a system error message and invalidating the measurements from various sensors.
  • Corrected minor user interface refresh errors.

Back to top of page

R10F

EVM [All Models]

Updates

  • Fully display the toxic serial number in the UNIT INFO screen (the last digit was truncated).
  • Large log files can now be downloaded without the system resetting.
  • The Particulate Matter (PM) user correction factor is now immediately applied (a power cycle used to be required).

Back to top of page

R10E

EVM [All Models]

Product Launch

  • First official firmware release.

Back to top of page