NoisePro®
Series Firmware Releases:
Release Level |
Date |
Purpose |
Type |
| 1.72 |
04/16/2006 |
MSHA |
N/A |
| 1.71 |
06/19/2005 |
Update |
Update |
| 1.70 |
06/05/2005 |
Update |
Update |
| 1.60 |
09/28/2004 |
Update |
Update |
| 1.58 |
08/09/2004 |
Update |
Update |
| 1.57 |
02/23/2004 |
Update |
Update |
| 1.56 |
N/A |
Internal Release |
N/A |
| 1.55 |
01/30/2004 |
Update |
Update |
| 1.54 |
12/18/2003 |
Update |
Update |
| 1.53 |
|
Update |
Update |
| 1.52 |
12/03/2003 |
Update |
Update |
| 1.51 |
|
Product Launch |
Product Launch |
|
|
| |
1.72 Custom MSHA modification
Custom modification for MSHA.
Update only required if customer is MSHA.
Back to top of page |
|
| |
1.71 Update
NoisePro® [All Models]
Updates
- Added serial communications commands to lock user setups to
these configurations or any other.
- Brought language translations up to date.
- Corrected "run flag" in firmware that was mapped to the same
location as "custom name 1".
- Added a 20 second delay after power-up whenever the unit is set
for Impulse time response. This allows time for the detector to
decay to ambient noise levels before collecting data. Display
indicates "IMPULSE RESPONSE SETUP" and counts down to zero from 20
while this is happening. Customers using the auto-on feature and
Impulse time constant will experience a 20 second delayed start in
data collection.
Back to top of page |
|
| |
1.70 Update
NoisePro® [All Models]
Back to top of page
|
|
| |
1.60 Update
NoisePro® [All Models]
- Displayed Data Errors
[Severity: Medium]
Values for TWA and SEL displayed on the instrument display may be in
error from few tenths to 1.0dB. The actual data dowloaded to QSP is
correct.
Back to top of page
|
|
| |
1.58 Update
NoisePro® [All Models]
- Battery Removal Issue
[Severity: High]
Accomplishes orderly shutdown of unit if the battery power is removed
during a study. Insures that only the current running study is lost and
all previous studies remain intact.
- Low Battery Issue
[Severity: Medium]
Prevents studies from starting when in a low battery condition. Also,
adds indicators in QSP and printouts to indicate if unit turns off due
to low battery condition.
- Microphone Removal Detection
[Severity: Medium]
Detects removal of microphone during a study and indicates this event in
both QSP and printouts.
- Studies with a Length of Zero Time
[Severity: Low]
Prevents user from creating sessions or studies that have a duration of
zero time.
Back to top of page
|
|
| |
1.57 Update
- NoisePro DL®
[Severity: None]
Logging Slow Max when Impulse time constant selected. Should be disabled
as modes incompatible.
- NoisePro® [All Models]
[Severity: Low]
Corrected issues with printing reports when set for Impulse mode. Did
not print Time Constant and printed Fast Max, not Impulse Max.
Back to top of page
|
|
| |
1.56 Update
- Internal Release
[Severity: None]
Back to top of page
|
|
| |
1.55 Update
- Calibration Problem
[Severity: Medium]
Corrected a pre and post calibration problem. If the instrument is set
in the low range and then calibrated the unit would give a low
calibration reading due to the unit calculating the SPL value as if it
were set to the high range.
- Verbiage Error
[Severity: Low]
Changed verbiage in the Italian calibration screens.
Back to top of page
|
|
| |
1.54 Update
- NoisePro DL® Problem
[Severity: Low]
Corrected issue with the NoisePro DL [ONLY] which logged two dosimeters
even if only one enabled.
- Enhancements:
[Severity: None]
Display shows current study number.
Printout shows dosimeter model.
Back to top of page
|
|
| |
1.53 Update
- Auto On Problem
[Severity: High]
Corrected unreliable Auto-On function. It would turn on but not
necessarily go into Run mode.
- Lavg/Leq Problem
[Severity: High]
Corrected a problem where Lavg or Leq was sometimes logged at twice it’s
value. This would be seen in a 5dB or 3dB increase respectively when
logging high values or using longer logging intervals.
- QuestSuite Professional Problem
[Severity: Medium]
Corrected a problem when using QuestSuite Professional to set up the
NoisePro. QSP did not reliably set up the dosimeters range and
weighting.
Back to top of page
|
|
| |
1.52 Update
- Low Battery Shutdown Issue
[Severity: High]
Corrected Low battery shut down to assure proper data storage.
- Min/Max/Peak Issue
[Severity: Medium]
Corrected a problem where incorrect session Minimums, Maximums, and
Peaks may be displayed if the unit was turned off between sessions.
- Printout problem with 24 hour Sessions
[Severity: Medium]
Corrected a printout problem related to end time calculations for
Sessions over 24 hours.
- Post Calibration Issue
[Severity: Low]
Corrected the post calibration display values.
Back to top of page
|
|
|
 |
|